When a bone is fractured, bone-plate is inserted into the human body to fix the bone. Usually the plate is made of metallic materials proved biocompatible safety. In the environment where bone begins to be recovered just, it is disadvantageous to act for the simple high strength plate. The reason is that this sites do not provide the best environment where according to the study of Perren, the initial resilience is the best when inter-fragmentary strain of the bone is 2~10%. In this study, Compliant Mechanism was applied to the Locking Compression Plate in order to design a plate with improved the drawback. Then, it was analyzed using a finite element analysis program ABAQUS and compared to the conventional bone-plate.