Трехмерная верификация дозиметрических планов при радиохирургии злокачественных новообразований в легком

O. Yu. Anikeeva, Пашковская Оксана Александровна, Филатов Петр Валерьевич et al. · JOURNAL of SIBERIAN MEDICAL SCIENCES · 2012

Assumed plans of 10 patients, who have been performed the pulmonary cancer and metastasises treatment, were investigated in the work. Verification of the medical plans performed with rotation irradiation modulation activity technique (VMAT). Treatment was carried out in one-stage and three-stage radiation mode with 15-20 Hz dose for fraction. Procedures were performed on the Elekta Axesse accelerator equipped with the EPID device.Verification of plans was carried out with EPID device and DosimetryCheck software. The average deviation of dose for the assumed and performed plans in the isocenter made 1,23 %. Three-dimensional γ-analysis showed γmid = 0,3 ± 0,02, γmax = 1,10 ± 0,12 and PTV volume with γ ≤ 1 equal to the 97,95 %.