Сильноточный импульсно-периодический ускоритель электронов прямого действия как средство стерилизации медицинских изделий однократного применения

Ростов Владислав Владимирович, П. И. Алексеенко, Выходцев Павел Васильевич et al. · Известия Томского политехнического университета. Инжиниринг георесурсов · 2012

The object investigated by the authors is searching for the safe and economical graphite-based adsorbent dressing sterilization method. The created high-current impulse-periodic electronic accelerator of direct action with accelerated electron beam lead-out in the atmosphere allows sterilizing the new adsorbent dressing. Material and analysis: bacteriological studies have been used to check the sterility of the dressing; receptivity and adsorptivity of the adsorbent dressing before and after sterilization, the influence of sterilization to receptivity and adsorptivity have been also studied. Conclusion: to guarantee the full graphite-based adsorbent dressing sterilization it is necessary to use not less than 25 impulses of electron beam raying. The summarized absorbed dozes are: D=17 kGy for a Spanlase-based dressing, D=13 kGy for a dressing based on decalcified filter. The both dozes are lower than analogous ones in application of γ-raying bandage sterilization. Receptivity and adsorptivity of the adsorbent dressing after sterilization have not changed.

Cited by 0 works.