족관절 보조기를 착용한 젊은 사람과 고령자의 족저굴국 운동 토크 보조 특성 비교

김경, 강승록, 박용군 et al. · 제어로봇시스템학회 합동학술대회 논문집 · 2009

Ankle-foot orthosis(AFO) with a pneumatic rubber actuator, which is intended for the assistance and the enhancement of ankle muscular activities was developed. In this study, appropriate amount of power pattern of the device and the effectiveness of the system was investigated during plantarflexion motion of ankle joint. To find the appropriate amount of power pattern and the effectiveness of the system, the subjects performed maximal voluntary isokinetic plantarflexion contraction of a Biodex-dynamometer. Plantarfexion torque of the ankle joint is assisted by subject’s soleus muscle that is generated when ankle joint do plantarflexion motion. We compared the maximal plantarflexion torque between with active control of the AFO and without it in each ten young and elderly adults. The subjects were performed the gait motion with and without the active control using muscle. We made a comparison the muscular activities between with and without the active control. The experiment results in plantarflexion motion showed that the muscular activities and peak torques with the AFO were reduced and the one wearing it were increased. Therefore, we confirmed the effectiveness of the edveloped AFO.