구강 내부 온도 계측을 위한 센서 시스템 연구
金敬昊 · 2007
In this study, a novel sensor system for measuring the temperature inside an oral cavity is proposed. With this aim, a small size of thermistor was used for resolving the cavity's temperature with the resolution of 0.1℃. To evaluate effectiveness of our sensor system, the temperature and its output voltage characteristic, and the specifications of response are investigated. It turned out to be that our sensor system has a linear property in terms of temperature variations for a healthy subject's body temperature range and has a good response time within 3 seconds. Also, in order to investigate the medical application, our sensor system is sought to measure the real temperature variations of a subject's oral cavity and ark shell especially for 'before' and 'after' exercise mode.
Cited by 0 works.