Токсикологическая оценка использования на крысах локального ультразвукового воздействия средней интенсивности, генерируемого установкой «УСДТ»

Бухман Владимир Михайлович, I. A. Merkulova, Н. П. Ермакова et al. · Российский биотерапевтический журнал · 2011

Potential safety for and the personnel of the ultrasonic influence which is carried out by means of the new domestic device for carrying out sonodynamic therapy “USDT” is estimated in experiments on males adult outbred rats. The site of a rat’s tail (the area 0,9 мм 2, 3-5 cm from the tai l basis) was exposed to ultrasound generati ng from two transducers by frequency of 0.7 and 2.64 MHz with intensity from 3 to 6 W/cm 2 for 10 minutes. The unitary scanning, triple scanning and unitary directed regimens of an irradiation in doses from 13.5 to 81.0 kJ were used. We have not observed influences on an organism of animals of group personnel. Concerning rats of group patients it is not taped side effects when we used therapeutic regimen of 3 W/cm 2 regimen. The all-toxic actions were not observed in used regimens. Local toxicity developed at intensity >3 W/cm 2 in scanning regimen and in all interval of doses at unitary directed regimen. The data obtained indicate that the application of device “USDT” is safety for and the personnel when used at a therapeutic scanning regimen.