Система формирования пучка комплекса протонной терапии ияи РАН
Sergey V. Akulinichev, Veselin Vasilev, Ю. К. Гаврилов et al. · CyberLeninK (CyberLeninka) · 2008
A new original beam formation system for proton therapy was developed and built in INR of RAS. This system, based on the passive beam formation method, consists of a automatic set of collimators, energy degrader with a variable water lay, dynamic beam modulator, emergency and routine beam shutters, brash filters, boluses etc. The position of all the elements of this system is realized with the accuracy better than 0.1 mm. The formation system provides the irradiation of any targets from eye tumors of few millimeters up to deep seated tumors with the size of 100 mm. The deviation of the delivered dose from the plan should not exceed 5%.
Cited by 0 works.