정밀온도측정을 위한 서미스터 교정
감기술, 김용규, 양인석 · 2011
We demonstrated that high-stability thermistors can be calibrated with an uncertainty less than 1 mK, if the error due to the heat conduction is minimized. We first investigated the effect of the self-heating of typical thermistor probes to see how accurate we need to determine the effect of self-heating. We, then, calibrated thermistors and fitted the results using various modeling equations. We found out that the heat conduction is an important factor in achieving the calibration uncertainty under 1 mK for thermistors when the diameter of the probe is as thick as 10 mm. Therefore, we controlled the room temperature within 0.5 。C to minimize the heat conduction error during the calibration. The calibration with an uncertainty below 1 mK was possible when the stabilization time for each calibration was long enough to obtain a good thermal equilibrium.