Валидационные исследования ВЭЖХ-методики анализа экспериментального препарата «Премелтоп»
В. М. Воробьева, О. Г. Макарова · Современные проблемы науки и образования · 2013
Validation HPLC-methods of qualitative and quantitative analysis of metronidazole, lidocaine hydrochloride, prednisolone in the structure of the experimental drug, «Premeltop» was performed. In liquid chromatograph LC-20 of the company «SHIMADZU» (Japan) with UV detector with further computer processing of the results of research using the program «LCsolution version 1.4» studies were conducted. Stationary phase chromatography column 4,6 x 150 mm PerfectChrom C18, a particle size of 5 nm («MZ-Analysentechnik», Germany). The mobile phase: trifluoroacetic acid a solution of 0.1%, B: acetonitrile 100%. Gradient elution concentration changes of eluent B from 10% to 80%, duration of the analysis 12 min. The technique for quality assessment Premeltop (a powder for oral suspension) can be used, as experimentally defined settings linearity (R2 = 0,9991; 0,9996; 0,9992), application range (50 to 150%), precision in terms of the frequency of occurrence (RSD=1,61%; 1,67%; 1,75%), correctness (=101,41±1,10%; 100,90±1,00%; 100,12±0,95%) for metronidazole, lidocaine hydrochloride, prednisolone, respectively, which do not exceed the recommended values.