Влияние параметров вакуумной системы течеискателя на характер изменения давления пробного газа в масс-спектрометрическом анализаторе

V. M. Myasnikov, Sergei V. Vinogradov, А. Н. Лобаев et al. · Современные проблемы науки и образования · 2012

Using the balance of trial gas streams arriving at the vacuum system of mass spectrometer leak detector and a stream formed as a result of roll-out by a fore-pump a mathematical model describing the process of mass transfer of trial gas in vacuum system of leak detector is received. Dependence of change in trial gas pressure in the analyzer of mass spectrometer leak detector has been defined for schemes of forward flow and backflow. A turbo-molecular pump with changing frequency of cylinder barrel rotation is used in vacuum system of leak detector. Possibility of operative definition of direct and return conductivity of a turbo-molecular pump and the speed of a fore-pump action are established under characteristics of the transients describing the change of partial pressure in the analyzer using the schemes of forward flow and backflow.

Cited by 0 works.

Authors