离子色谱-质谱联用法同时测定饮用水中3种痕量-氯酚

金米聪, 颜勇卿, 陈晓红 et al. · 2006

A novel, sensitive and accurate method has been successfully developed to the simultaneous determination of 3 monochlorophenols (MCP) in drinking water samples by ion chromatography (IC) coupled with an atmospheric pressure chemical ionization mass spectrometry in the negative mode. Drinking water was preconcentrated by solid-phase extraction (SPE). The IC separation was carried out with an IonPac(superscript ®) AS18 analytical column (250mm×4.0mm) and an IonPac(superscript ®) AG18 guard column (250mm×4.0mm) using 15mmoL/L KOH containing 10% acetonitrile as mobile phase in isocratic mode at a constant flow rate of 1.0mlJmin. The molecular ion m/z 127 [M-H](superscript -) was selected for the quantification in selected ion monitoring (SIM) mode. The relative standard deviations of intra-day and inter-day were less than 8.6% and 10.5%, respectively. The limits of quantification (LOQ) in water samples were 0.02, 0.01 and 0.01g/L for 2-MCP, 3-MCP and 4-MCP, respectively. The method allows the 3 target compounds in water samples to be determined at μg/L levels, and it has been successfully applied to the routine analysis.

Cited by 0 works.