회전근개 봉합술에서 나사못의 삽입각도가 고정력에 미치는 영향 분석
남승현, 박원만, 최대경 et al. · 대한기계학회 춘추학술대회 · 2014
Insertion angle of the anchor is an important factor for successful rotator cuff surgery. In spite of a number of experimental and computational studies on the insertion angle for increasing fixation power of the anchor, optimal insertion angle of the anchor is still an issue in orthopaedic medical engineering field. In this study, we investigated the effects of insertion angle of the anchor on fixation power in rotator cuff surgery by using finite element analysis. The anchor was inserted at 5mm posterior to bicipital groove with insertion angles of 45°, 67.5°, 90°, 135° from surface of humerus. A force of 217N was applied to the anchor with six different angles (15°, 30°, 45°, 60°, 75° and 90°) against the superior surface of humerus. Stiffness of the anchor after fixation and von-Mises stress on the bone around the anchor were analyzed in each loading condition. The results of this study showed that von-Mises stress and stiffness are correlated with the angle of the applied force. The obtuse angle showed the higher stiffness and the angle closer to 90° showed higher stress. Insertion angle of 135° showed higher fixation power than other insertion angles. The results of this study could provide fundamental information for developing optimal designs of the anchor and surgical methods in order to enhance rotator cuff repair.