高效液相色譜-在線消解-氫化物發生原子吸收光譜聯用技術研究

劉華琳, 趙蕊, 韋超 et al. · 2005

The coupling of high performance liquid chromatography (HPLC) with atomic absorption spectrometry (AAS) was achieved by using a simple laboratory-built ultraviolet (UV)-hydride generation (HG) interface. Compared with the former techinque, a 0.4 m quartz tube was used for UV digestion instead of the 5 m teflon coil without decreasing the UV digestion efficiency of the organic arsenic compounds. As a nesult, the volume of the interface was greatly minished, so the miniaturization and integration of the instrument was achieved. The coupling instrument was used to separate four arsenic species (arsenite, arsenate, methylar-sonic acid, dimethylarsinic acid) that are of toxic importance. The optimized analysis conditions were obtained. The results showed good repeatability (RSD>6.56%), good linearity (r>0.997) and low detection (detection limit>14.4 mg/L). Then the performance for on-line digestion of organic arsenic species including arsenobetaine (AsB) and arsenocholine (AsC) was investigated. A rapid, direct, continuous, online HPLC-UV-HGAAS hyphenated technique was developed.

Cited by 0 works.