체력 측정과 운동처방을 위한 통합형 체력검사 기기의 신뢰도 검증

김형돈, 이재석, 길재호 · 2010

The purpose of this study was to compare the measurement values obtained using integrated type fitness testing instrument and individual testing instruments and determine the reliability of integrated type fitness testing instrument. Sixty-six subjects consisting of 38 males and 28 females were volunteered to participate for the study. The subjects were tested for weight, percent body fat, grip strength, muscle endurance, agility, power, and cardiorespiratory endurance with two types of testing instrument separately. All subjects were undertaken exercise tests with the both of integrated instrument and individual instrument tests at an interval of less than seven days. Percent body fat was measured using bioimpedance method for both test. VO(2)max was determined with GXT using bicycle ergometer tests. The work load increased 15watts in every minute for subjects to the exhaustion. VO(2)max from the integrated instrument was also estimated with regression equation programmed for heart rate of submaximal work load. The data was analyzed with descriptive statistics, t-test, intraclass ccorrelation coefficient, and Pearson correlation coefficient using SPSS. The results showed that there was no significant differences between two testing methods except percent body fat, right hand grip strength, and VO(2)max of male and female subjects. The Peason correlation coefficient revealed that significant relationships exist between the test scores of two different testing methods. The significant intraclass coefficient correlation reliability was also observed from integrate type fitness instrument.

Cited by 0 works.