고체상추출법과 HPLC/MSD/FLD를 이용한 수질중의 페놀 분석

이태준, 박근영, 표동진 · Analytical Science and Technology · 2015

An analytical method for determining phenol considered priority pollutants of the US EPA and precursor of toxic phenolic compounds by solid-phase extraction (SPE) and high performance liquid chromatographic systems (HPLC) equipped with fluorescence and mass selective detectors have been developed. The SPE process for sample preconcentration was performed on a commercially available Oasis HLB cartridge packed with polymeric sorbents. The effect of pH, elution solvent, and elution volume on the recoveries of the analytes were investigated with HPLC/FLD. Average recovery of >87.0% was achieved with 60 mg sorbents using 5 mL of methanol as an elution solvent at pH=3.

Cited by 0 works.

Authors