О клиническом опыте мониторинга артериального давления

V K Kumykov, P. M. Захохов, Zalina Khasanovna Abazova et al. · Вестник Российского университета дружбы народов. Серия: Медицина · 2007

The paper is devoted to the specific feature analysis of blood pressure measurements. Probable sources of errors leading to the distortion of blood pressure values are discussed. Thus it is necessary to remember that measurements have to be done with empty bladder. Filled bladder squeezes thigh-artery and creates obstacles for blood circulation, resulting the changes of haemo-dynamic parameters, playing the main role in blood pressure regulation. Our measurements showed that bladder draining reduces the blood pressure on 10-15%. This may lead to incorrect interpretation of blood pressure values. During air release from the cuff it is necessary to take into consideration that the releasing procedure has to be not longer than 8-10 seconds. In other case the considerable deformation of blood vessels will course the increase of their hydrodynamic resistance and as a result the reduce of venous return and distorted values of blood pressure. During repeated blood pressure measurements the five minute pause is recommended for blood circulation recovery.

Cited by 0 works.

Authors