高效液相色谱-感耦合等离子体质谱联用技术测定二价汞、甲基汞、乙基汞与苯基汞
张兰, 陈玉红, 施燕支 et al. · 2009
A method was established for the determination of inorganic mercury, methyl-mercury, ethyl-mercury and phenyl-mercury by high performance liquid chromatography coupled with inductively coupled plasma-mass spectrometry (HPLC-ICP-MS). Ammonium acetate/L-cysteine/methanol were used as mobile phase for the separation of inorganic mercury, methyl-mercury, ethyl-mercury and phenyl-mercury by the gradient elution. The detect limits were 0.022 ng•ml^(-1) for inorganic mercury, 0.022 ng•ml^(-1) for methyl-mercury, 0.028 ng•ml^(-1) for ethyl-mercury and 0.041 ng•ml^(-1) for phenyl-mercury respectively. The relative standard deviations for peak area of 5.0 ng•ml^(-1) mixed standard solution were: 1.40% for inorganic mercury, 1.01% for methyl-mercury, 0.97% for ethyl-mercury, 2.13% for phenyl-mercury (n=6).
Cited by 0 works.