초기변형률에 의한 피부조직의 미소바늘관통저항 감소에 관한 연구
김종훈, 박성민, 김도훈 et al. · 대한기계학회 춘추학술대회 · 2015
Despite all the recent advances made in the biodegradable microneedle field, the mechanical failure of a microneedle during insertion remains a major technical hurdle in the path of large scale commercialization. A controllable reduction in the microneedle puncture resistance of skin tissues is therefore an essential requirement for developing a biodegradable microneedle. Here, we consider unidirectional or multidirectional prestress on skin tissues as a critical stimulus to reducing the microneedle puncture resistance, and investigate the effect of prestress on the changes in microneedle insertion behavior (e.g., puncture resistance, insertion force, etc.) in a quantitative way. For pork skin tissues similar to human ones, microneedle insertion behavior is characterized with a z-axis motorized stage equipped with a load cell, together with microindenters; the experimental apparatus provides a force-time curve which contains information on microneedle puncture and insertion into skin tissues. The experimental results lead to quantifying the relation between microneedle insertion behavior and prestrain. The findings of this study will lead to better understanding of skin tissue biomechanics.
Cited by 0 works.