유세포분석용 항인면역글로불린 반응성 비교 방법
원동일 · The Korean Journal of Laboratory Medicine · 2003
Background : In flow cytometry, the verification of the anti-human immunoglobulin (anti-lg) reactivity is important. The author has devised a simple protocol for this. Methods : As a staining verification control, red blood cells (RBCs) were used. IgG- or IgM- coated RBCs were made using group O or group B serum, respectively. Ani-Ig reactivity in each serially diluted serum for immunoglobulin coating and correlation between the amount of reacting anti-Ig on RBCs and measured values were investigated. In one case of parallel tests for two anti-Ig reagents, author's method was compared with the B cell gating method. Results : Both the test/control mean channel fluorescence ratio and percent fluorescence had significant linear correlations with the amount of the reacting anti-IgG on RBCs. As well as B cell gating, the author’s method seemed to also clearly discriminate the reactivity of two anti-IgG reagents. Conclusions : Using immunoglobulin-coated RBCs as a staining verification control, the parallel test for two anti-IgG reagents could be performed more conveniently and objectively than the B cell gating method.
Cited by 1 works.