“氣管插管鑒定囊”在急診搶救插管中的應用-與聽診法對導管位置判定的比較
童幼良, 孫鳴, 夏江燕 et al. · 2003
Objective To compare the efficacy of tracheal detecting-bulb (TDB) and stethoscope in critically ill patients. Methods Fifty-seven patients with cardiac arrest were enrolled. After tracheal intubation and ETT cuff inflation in the emergency department, the patients were allocated randomly to two groups. In group 1 (n=30), patients underwent a TDB test and an auscultation test in sequence. In group2 (n=27), patients underwent an auscultation test and a TDB test in sequence. Results The TDB test was faster than the auscultation test; The TDB test and auscultation test indicated all fourteen esophageal intubations. In 57 tracheal intubations, the TDB test correctly indicated 53 tracheal intubations (sensitivity, 92.98%). The ausculation test identified 46 tracheal intubations (sensitivity, 85.19%). Conclusion The TDB is a valuable diagnostic tool for differentiating tracheal from esophageal intubation.
Cited by 0 works.