온,습도 시험을 통한 자가 혈당 측정 센서의 안정성 평가
김선진, 함중걸, 박호준 et al. · 대한설비관리학회지 · 2013
A self-monitoring blood glucose sensor is a kind of in vitro diagnostic medical devices which converts electrical signal to glucose. Performance and stability testing are conducted primarily to evaluate glucose sensors. Stability testing determines shelf-life of glucose sensors, which is mainly performed accelerated testing. Previous stability testing has been done only using temperature condition. In this study, we add humidity condition in order to reduce testing time. As a result, in the case of glucose sensors of manufacturer A, we can show accelerating temperature and humidity was degradated faster than accelerating temperature only. On the other hand, however, glucose sensors of manufacturer B degradated rapidly in temperature condition. We found that degradation pattern changes, depending on the type of glucose sensors` enzyme and intermediate.
Cited by 0 works.