광센서를 이용한 실시간 중환자 요량감시 장치

김종명, 이진영, 홍주현 et al. · 센서학회지 = Journal of the Korean Sensors Society · 2008

【This paper addressed real-time urine monitoring device for intensive care patients. The device was developed to detect and count each urine drop using optical sensor and calculate the current urine output volume and its hourly rate. In experiment, the water volume scale of drainage bottle was observed and compared with the count of the device so that the volume of each drop was found to vary with the dropping rate per minute. From this measurement, the relationship equation was derived to estimate the total water volume from the drop rate (correlation coefficient : r= 0.99). The developed device could be applied to count patient's urine drop successfully. Therefore, this device can be used to monitor intensive care patient's urine status in real-time.】

Cited by 0 works.