무구속 건강모니터링을 위한 심탄도 계측 시스템 구현 및 평가

노윤홍, 정도운 · 센서학회지 = Journal of the Korean Sensors Society · 2010

【This research proposes measuring of BCG(ballistocardiogram) to monitor heart activities in a non-constrained environment, at home or work. Unlike with ECG, measuring BCG does not require the attachment of leads on the subject's body and allows signal measuring in a non-constrained state. It enables effective long-term monitoring of cardiac conditions. In this study a chair type BCG measurement system to continuous monitor the activity of the heart is implemented. The instrument consists of upper petal and ready for press of chair load cell sensor is attached to measure the change of the object's weight. In order to extract the output ballistic signal from the weight and force sensor signals. Beside the signal processing circuit for the digital conversion, the ballistic signal is detected using DAQ equipment. Signal processing algorithm including wavelet transforms for noise cancellation, template matching for normalization and peak detection in BCG is developed. ECG and BCG were concurrently measured to evaluate the performance of the system, and comparing the characteristics of the two signals verified the possibility of the system in non-constrained and nonconscious health monitoring.】

Cited by 0 works.