양전자 방출 단층 영상장치의 시스템 구조 및 촬영대상에 특화된 반복적 사이노그램 공백 보간 방법

김광돈, 이기성, 이학재 et al. · 대한전자공학회 학술대회 · 2016

Various geometries have been used to develop the positron emission tomography (PET) scanner in order to improve PET image resolution, and to detect events with higher sensitivity. However, the blind zones or inter-detector gaps for each detector block result in various artifacts and image distortions in reconstruction images. To compensate for this distortion, the literature offers several gap-filling methods, such as linear interpolation methods and iterative gap-filling methods. Uygar Tuna et al. proposed a novel gap-filling algorithm that shows superior gap-filling results, which can be attributed to the PET scanner geometry. However, the algorithm is not dedicated to imaging objects, and should be manually optimized for these objects. We aim to develop a gap-filling algorithm, in particular the filter mask design method of the algorithm, which optimizes the filter to the imaging object by an adaptive and iterative process, rather than by manual means. We generated the data set with two analytic phantom studies, Shepp-Logan and Jaszczak. The proposed algorithm shows results that compare to those of the manually optimized discrete-cosine transform 2 (DCT2) algorithm without perfect or full information of the imaging object, and improve the image quality of the reconstructed image.

Cited by 0 works.