혈액중 포도당과 젖산의 분석을 위한 광섬유 생물센서
손옥재, 이종일 · 센서학회지 = Journal of the Korean Sensors Society · 2006
【Optical-fiber sensors have been developed to determine the concentrations of glucose and lactic acid in blood samples. Fluorescence dye [tris(2,2'-biphenyridine)-ruthenium(II)-chloride (RuBPY)] was entrapped by using a silicon to the unclad tip of a glass optic fiber. Enzymes like glucose oxidase (GOD) and lactate oxidase (LOD) have been immobilized by acrylamide resin adhesive, adsorption with zeolite or covalent bonding with aminopropyl-triethoxysilan. The fiber-optic glucose/lactate sensor was then used to analyze the concentrations of glucose and lactate in blood samples. The results were compared with the results of HPLC analysis and their difference was in error by less then 5 %.】
Cited by 0 works.