An improved filtered back-projection algorithm using pre-processing
Thomas J. Hebert, S.S. Gopal · IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) · 2002
The theoretical basis for the filtered backprojection (FBP) algorithm is analyzed, and the underlying assumptions inherent in its use in single photon emission computed tomography (SPECT) are examined. Using both theoretical considerations and experimental investigation, the authors propose an improvement to the FBP algorithm for application to SPECT imaging. In the proposed modification to the filtered backprojection algorithm, the projection images are preprocessed so that the SPECT projection images more closely fit the model by which the FBP algorithm is formulated. The authors compare this approach with standard FBP in SPECT phantom and patient studies.>
Cited by 12 works. doi:10.1109/nssmic.1991.259279
Authors
References
- Maximum Likelihood Reconstruction for Emission Tomography
- The Fourier reconstruction of a head section
- Topics in Applied Physics
- Image reconstruction from projections
- A Maximum Likelihood Approach to Emission Image Reconstruction from Projections
- Effects of myocardial wall thickness on SPECT quantification
- Attenuation Correction and Incomplete Projection in Single Photon Emission Computed Tomography
- A projector, backprojector pair which accounts for the two-dimensional depth and distance dependent blurring in SPECT
- A comparison of deconvolution and windowed subtraction techniques for scatter compensation in SPECT
- Statistical stopping criteria for iterative maximum likelihood reconstruction of emission images