Накостный остеосинтез переломов нижней трети диафиза плечевой кости

Igor' Ivanovich Litvinov, В. В. Ключевский, Александр Александрович Рыжкин · Traumatology and Orthopedics of Russia · 2011

On the basis of three-dimensional analysis of the humerus bones was shown that the plate, placed on posterior-and-medial surface of the lower section of these bones, should be twisted in terminal portions and bended on both ends in the contact surface direction and should have an intermediate untwist sector. The results among 31 operatively treated patients with 32 distal third diaphyseal fractures of the humerus were analyzed. Was stated that the modified osteosynthesis of distal third diaphyseal fractures of the humerus associates with lower risk of iatrogenic damage it, provides fixation stability and significantly better outcomes (Students criterion – 2,7), which allows to use that method of internal fixation in cases where the performance of intramedullary nailig is not possible.

Cited by 1 works. doi:10.21823/2311-2905-2011-0-1-117-120 Open-access PDF