Моделирование свободных колебаний звукопроводящей системы реконструированного среднего уха

Gennadi I. Mikhasev, М. А. Фирсов, В. П. Ситников · Российский журнал биомеханики · 2005

The simple mathematical and mechanical model of the reconstructed oscillating system of a human middle ear has been presented. The reconstructed oscillating system contains a round plate made of cartilage, and two bars hinged one to another. The first bar is rigidly connected with the artificial tympanic membrane, and it models a T-type prosthesis that substitutes the chain of auditory ossicles hammer-anvil, the second bar models the stirrup ossicle. There has been derived a set of simultaneous non-linear differential equations that describes the free radially non-symmetric vibrations of the middle ear mechanical system. The transcendental equation with respect to natural frequencies of the system has been derived for the case of linear vibrations. The effects of mechanical and physical characteristics of the introduced prosthesis on the lower natural frequencies spectrum are under investigation. The frequencies of the reconstructed oscillating system are compared with the data by other authors on natural frequencies of normal human middle ear that were obtained by the finite element modelling.

Cited by 0 works.

Authors