КОМПЛЕКС ДЛЯ ЛОКАЛЬНОЙ ИОННОЙ ИМПЛАНТАЦИИ И ЭЛЕМЕНТНОГО МИКРОАНАЛИЗА С ВЫВОДОМ ИОННОГО ПУЧКА В АТМОСФЕРУ
Fadei Fadeevich Komarov, А. С. Камышан, Vladimir V. Pilko et al. · Doklady of the National Academy of Sciences of Belarus · 2016
A measuring system for local elemental analysis is constructed on the basis of an ion accelerator AN-2500. It comprises a device of an in-air ion beam extraction through a tapered glass capillary and a PIXE spectrometer. The angular dependence of the proton beam possessing the transmission energy of 150-250 keV through the tapered capillary with the output diameters of 5 and 10μm is determined. The transit of protons through the capillary is found to be in threshold character and to be determined by the charging rate of its internal surface. The characteristics of current and energy of the ion beam extracted into the atmosphere are analyzed. It is found that the value of output current can be varied from 0.5 to 3 nA, while the ion energy is almost equal to the initial beam energy. The PIXE energy spectra of various materials are measured at the atmospheric pressure. It is found that the output beam intensity is enough for an accurate determination of the elemental composition of the samples under analysis.
Cited by 0 works.