МАЛОГАБАРИТНЫЙ ЯДЕРНО-МАГНИТНЫЙ СПЕКТРОМЕТР ДЛЯ ЭКСПРЕСС-АНАЛИЗА ЖИДКИХ СРЕД
Давыдов Вадим Владимирович, A. Yu. Karseev · CyberLeninK (CyberLeninka) · 2013
Design of the handheld nuclear – magnetic spectrometer for the measuring of spontaneous T 1 and transverse T 2 relaxation times of the liquid medium in low magnetic field B 0 = 55 mTl is developed. Julotto method for the measuring of T 1 in developed spectrometer design in low magnetic field is implemented for the first time. This method was previously applied only for measurement in strong magnetic field for solid samples in which the ratio of the relaxation times is different from liquid mediums. The nuclear magnetic resonance signal acquisition scheme is developed and realized. This makes it possible to carry out relaxation times measuring for small amount of the liquid medium (10 – 15 gram) when ratio of signal/noise is about 1.5.
Cited by 0 works.