혈액 분석기기의 신뢰도 및 타당도 분석
엄한주, 강성기 · 한국체육학회지 · 2004
This study investigates the reliability and validity of blood lipid analyzers. Four blood analyzers, two different types (DT 60, Hitachi 7600-110,) with 2 instruments for each type, were used to compute the values of TC, HDL-C, and TG. Also, Lactic acid concentrations were repeatedly obtained by two instruments (DT 60 and Cobas integra, each type having 2 instruments). In all analyses, the same blood samples from each of 6 subjects were used to examine the repeatability and reproducibility of the values measured by all instruments. Data were analyzed to compute coefficient of variations (CV), variance components, correlations and various descriptive statistics with SPSS 11.0. HDL-C and TC values obtained by the same type of analyser for the 3 repeated samples from the same person under each condition were highly consistent, the CV for each being within 5%. However, when the values obtained from two different types of the analyzers were combined, the CV increased up to 12.3% for HDL-C and 21.7 ~ 24.5% for TC, indicating very low reproducibility of the analyzers. The correlations between the values of each measured variables with the same type of analyzers were very high, reaching above .97, but those values obtained by different types of analyzers were considerably lower. The analyses of variance component also revealed that the main source of error variances were mostly due to the difference in the types of analyzer. Thus, these results suggest that the different types of analyzers do not produce the same results although the 2 types are designed and manufactured to have the same function in analyzing blood lipid. For the TG and Lactic acid variables, the results indicated that there was insufficient evidence of both repeatability and reproducibility of the analyzers as the values obtained repeatedly from the same type of analyzers as well as those obtained from different types of analyzers were inconsistent. In general, the blood analyzers for TC and HDL-C were reliable, but not valid. However, the values of TG and Lactic acid were neither reliable nor valid as the values were varied considerably both within the same type and between the different types of analyzers. HDL-C and TC concentration is within coefficient of variation(CV) 5%. A similar Hitachi 7600-110 and DT 60 blood analyzer whole mean concentration of participants lower. However, defined as the calculated HDL-C (CV:12.8-16.36%), TC (CV: 21.68-24.47%), TC concentration should alter with change the analyzer to estimating blood range. Both HDL-C and TC concentration is above correlate .94 respectively. Therefor, to analyse HDL-C and TC, we found high mean different. Variance component estimation of HDL-C and TC are HDL-C(33.88%), TC(11.65%) which is different participant and another variance component estimation is different blood analyzer(HDL-C 6205%, TC 86.11%). To analyse TG and Lactic acid, we found estimation values of blood analyzer. Variance component estimation of TG and Lactic acid are TG(627%), Lactic add(87.9%) of different blood analyzer.
Cited by 0 works.