Десятилетний опыт использования отечественного искусственного клапана сердца «МедИнж» при митральном протезировании
Немченко Евгений Владимирович, С С Степанов, В. К. Новиков · Вестник Санкт-Петербургского университета. Серия 11. Медицина · 2006
In this retrospective study, between April 1995 and April 2005 233 patients receivedmitral valve replacements with «MEDING» prothesis and in control group 12 «SaintJude Medical», 20 «ATS Medical» and 13 «Carbomedics» valves were implanted. CABG and aortic valve replacement were excluded from this study. Follow-up was made in 55,8% patients (59,5% in control group) with average term3,5±2,8 years, 4,5±2,4 years for controlgroup; p>0,05. Hospital mortality was 3% and peak gradient on the prosthesis was 1 l,3l3,4 mm Hg (6,7% and 10,0±4,4 mm Hg in control group; p>0,05). Survival rate, freedomfromnon-fatalthromboembolismand reoperationwere8l,8%, 86,3 and 94,4% after 5 year follow-up, in 9 years 67,8%, 75,1 and 91%, accordingfy Comparison of actuary curves did not reveal any distinctions from control group (p>0,05). «MEDING» prosthetic heart valve provides good early and late results which are similar to «Saint Jude Medical», «ATS Medical» and «Carbomedics» implantation. Keywords: prosthetic heart valve, MEDING, mitral valve replacement, results of implantation.