Метрологічний аналіз та метрологічне забезпечення атомно-силової мікроскопії
Х.О. Надорожняк, Л. А. Витвицька, Оксана Степанівна Литвин et al. · METHODS AND DEVICES OF QUALITY CONTROL · 2011
35 Grounded necessity of development and introduction of standards in relation to the metrology providing of sweepable probe microscopes as most perspective facilities of measuring technique in the area of nanotechnology and their using for the certification,, inspection proof-testing. The metrology analysis of atomic-power microscope of NanoScope IIIa Dimension 3000 is conducted, his metrology model and method and facilities of his calibration is developed. On the basis of experimental information variations of values of geometrical parameters of surfaces are determined in the nanometric range of 3D-measurement, caused by influence of different sources of uncertainty. The value of total uncertainty is expected, which is 0,6746 nm, that validifies and exactness of the three-dimensional drawing a map of surface by an atomic-power microscope
Cited by 0 works. Open-access PDF