РОЗРОБКА ДВОПРОМЕНЕВОГО ВИМІРЮВАЧА КОНЦЕНТРАЦІЇ ТА ДИСПЕРСНОСТІ ВУГІЛЬНОГО ПИЛУ З КОМПЕНСАЦІЄЮ ТЕМПЕРАТУРНОГО ДРЕЙФУ
Роман Ігоревич Соломічев, Олександр Володимирович Вовна, А.А. Зори · Visnyk of Vinnytsia Politechnical Institute · 2014
Concentration and dispersion meter of coal dust is developed and studied with temperature drift compensation. In its la-boratory tests is set the basic absolute error of dust concentration measurement in the range from 0 to 3 g/m3, which does not exceed ±8 mg/m3, which is more than less a magnitude order of the dust concentration measurement absolute error of the aspiration measuring devices (±200 mg / m3). Additional dust concentration measurement error in the temperature range from 5 to +35 ○C is ±2,5 mg/m3 that no more than 35 % of the dust concentration measuring basic error and totally satisfies the meter operation conditions of the mine atmosphere.
Cited by 0 works.