Apossibility of in-process on-line monitoring of the agglomerate chemical composition is studied using a CON-X 02 pipeline x-ray fluorecence analyzer. Relative error of Fe determination in the chemical composition of sinter is 0.30 % rel. at a content of 56 %; 2.33 % rel. for calcium oxide at a content of 6%; and 6.33 % rel. for basicity at a value of1.45 units. We proposed and experimentally tested a model for calculating chemical composition of the agglomerate proceeding from the chemical composition of the components and their consumption for sinter production.