원저 : 기관내삽관 확인방법으로 일회용 호기말 이산화탄소 측정기의 이용

오동렬, 박규남, 이원재 et al. · 대한응급의학회지 · 1994

Study objective:. To determine the ability of a disposable end-tidal CO detector to accurately confirm or refute endotracheal tube placement. Design.: Seventy prospective emergency intubations. Setting: Emergency intubations performed in the emergency department. Type of participants: Intubations were performed by emergency medicine residents and interns. Interventions: The FEF CO, detector was applied after 70 emergency intubations. Notation of color change indicating intratracheal placement was recorded in each case. Confirrnation or refutation of the detectors results was determined subsequently through traditional methods. Results: The sensitivity for confimation of endotracheal intubation in 50 nonarrested patients was 100%. However, only 14 of 20 patients(sensitivity, 78% ) in cardiac arrest had endotracheal intubation in 20 arrested patients in whom color change was noted(100%). Overall sensitivi ty for tracheal intubation was 94%, and specificity for tracheal intubat,ion was 100%. Conclusion: The FEF CO, detector reliably detects intratracheal placement in t,he nonar- rested patient. Its use in prolonged cardiac arrest merits further study.