상지 자기공명영상 검사 시 호흡에 따른 운동 인공물 감소를 위한 보조 고정 기구의 유용성 평가
권순용, 김상우, 김성호 et al. · 2020
In this study, we aimed to reduce of motion artifact of the upper extremity caused by breathing by designing an immobilization device. The upper extremity was classified into three regions including upper arm, elbow, and forearm to observe the effect on the motion artifact. T1 and T2 weighted images(T1w, T2w) were analyzed through the mean signal intensity (MSI) Image analysis was achieved by measuring the average signal value in the background region in the phase code direction to evaluate the degree of motion artifact. In the result without using an immobilization device, the MSIs showed significant differences according to the anatomical regions(p<0.00). Especially, the MSI at the upper arm was increased by about 2.2~2.5 times compared to the forearm, while the elbow region was about 1.05~1.11 times that of the forearm. On the other hand, the use of the immobilization device resulted in substantial reductions of the MSIs at all anatomical regions(p<0.05). The MSI at T1w exhibited 62.68% at the upper arm, 39.78% at the elbo), and 47.03% at the forearm. In addition, those of T2w led to decreases in 58.22% at the upper arm, 38.22% at the elbow, and 46.49% at the forearm. The application of immobilization device to the extremity examination showed remarkable improvement of the motion artifact, particularly at the upper arm with a decrease in 58.22%~62.68%. Consequently, this study suggests that the immobilization device would be contributed to reducing the motion artifact, increasing the efficiency of the MRI scans and a more comfortable position for the patients taking an upper extremity MRI.
Cited by 0 works.