자기공명 두경부 혈관조영술에서 조영제 희석주입 비율에 따른 신호강도 분석

부재철, 강충환, 이정우 et al. · 2020

The present study analyzed the signal intensity differences depending on the diluted injection ratios of contrast medium (1.0 mol/l) and physiological saline when using time-resolved magnetic resonance angiography (DCE-MRA) for head and neck imaging. 3.0 T MRI was used as the research method, an automatic injector with the function of simultaneous injection of contrast medium and physiological saline was used, and dilution ratio (contrast medium: saline solution) compared were 10:0 (1000 mmol/l, undiluted solution), 7:3 (700 mmol/l), 5:5 (500 mmol/l) into three groups. For the injection method, the injection rate was 3.0 ㎖/sec on the right arm vein, and injection volume was 0.1 ㎖/㎏. Seven artery regions and two vein regions were selected as signal intensity measurement regions, while the measurement phase was set for dynamic multi-phase images of pure arterial phase prior to vein intrusion, the highest signal arterial phase, and the highest signal venous phase. The research results showed signal intensity increased in the ratio order of 10:0 (1000 mmol/l), 7:3 (700 mmol/l), 5:5 (500 mmol/l) for all measurement phases with the exception of transverse sinus for the highest signal arterial phase(p<0.01). Therefore, in DCE-MRA of head and neck imaging, the highest signal intensity can be achieved and optimal images obtained through 5:5 diluted injection of contrast medium and saline solution.

Cited by 0 works.

Authors