ДИФЕРЕНЦІАЛЬНИЙ ПРИЛАД ДЛЯ ВИМІРЮВАННЯ КОЕФІЦІЄНТУ ЕМІСІЇ ПОВЕРХНІ
Леонід Йосипович Воробйов, Леонід Декуша, Олег Декуша et al. · 2020
The urgency of creating a portable device for express-control of the emission coefficient (emissivity) of surfaces of materials and coatings is shown. The principle of operation and design of a differential device for measuring the emission coefficient are considered. The sensitive elements of the device are thermoelectric reclaims of heat flow and temperature. Computer simulation of complex radiation processes and convective-conductive heat transfer in cylindrical working cells of a portable device has been carried out. It was found that due to the differential scheme of the device, the spatial orientation of the cells practically does not affect the measurement results. The values of nonlinearity and the shift of the calibration characteristic due to convective heat transfer are estimated, which must be taken into account in the algorithm for processing information from the primary sensors of the device.
Cited by 0 works. doi:10.24027/2306-7039.3а.2020.218721
Authors
References
- Development of flight hardware for a large, inflatable-deployable antenna experiment
- Method for measurement of emissivity and absorptivity of highly reflective surfaces from 20 K to room temperatures
- TiO2-Siloxane Thermal Control Coatings for Protection of Spacecraft Polymers
- The Heat Exchange Simulation In The Device For Measuring The Emissivity Of Coatings And Material Surfaces
- Heat Exchange Simulation Of The Method And Portable Device For Measuring The Emissivity