Computed Tomography-Navigation™ Electromagnetic System Compared to Conventional Computed Tomography Guidance for Percutaneous Lung Biopsy: A Single-Center Experience
Morgane Lanouzière, Olivier Varbédian, Olivier Chevallier et al. · Diagnostics · 2021
The aim of our study was to assess the efficacy of a computed tomography (CT)-Navigation™ electromagnetic system compared to conventional CT methods for percutaneous lung biopsies (PLB). In this single-center retrospective study, data of a CT-Navigation™ system guided PLB (NAV-group) and conventional CT PLB (CT-group) performed between January 2017 and February 2020 were reviewed. The primary endpoint was the diagnostic success. Secondary endpoints were technical success, total procedure duration, number of CT acquisitions and the dose length product (DLP) during step ∆1 (from planning to initial needle placement), step ∆2 (progression to target), and the entire intervention (from planning to final control) and complications. Additional parameters were recorded, such as the lesion’s size and trajectory angles. Sixty patients were included in each group. The lesions median size and median values of the two trajectory angles were significantly lower (20 vs. 29.5 mm, p = 0.006) and higher in the NAV-group (15.5° and 10° vs. 6° and 1°; p < 0.01), respectively. Technical and diagnostic success rates were similar in both groups, respectively 95% and 93.3% in the NAV-group, and 93.3% and 91.6% in the CT-group. There was no significant difference in total procedure duration (p = 0.487) and total number of CT acquisitions (p = 0.066), but the DLP was significantly lower in the NAV-group (p < 0.01). There was no significant difference in complication rate. For PLB, CT-Navigation™ system is efficient and safe as compared to the conventional CT method.
Cited by 25 works. doi:10.3390/diagnostics11091532 Open-access PDF
Authors
- Morgane Lanouzière
- Olivier Varbédian
- Olivier Chevallier
- Loïc Griviau
- Kévin Guillen
- Romain Popoff
- Serge Aho
- Romaric Loffroy
References
- CT-guided needle biopsy of lung lesions: A survey of severe complication based on 9783 biopsies in Japan
- Proposal of a New Adverse Event Classification by the Society of Interventional Radiology Standards of Practice Committee
- Complication rates of CT-guided transthoracic lung biopsy: meta-analysis
- CT fluoroscopy-guided lung biopsy versus conventional CT-guided lung biopsy: a prospective controlled study to assess radiation doses and diagnostic performance
- Clinical Utility of Real-time Fusion Guidance for Biopsy and Ablation
- CT fluoroscopy-guided vs. multislice CT biopsy mode-guided lung biopsies: Accuracy, complications and radiation dose
- Advantages of Optical Compared with Electromagnetic Tracking
- Electromagnetic stereotactic navigation for external ventricular drain placement in the intensive care unit
- Computer assisted electromagnetic navigation improves accuracy in computed tomography guided interventions: A prospective randomized clinical trial
- Percutaneous lung biopsy: comparison between an augmented reality CT navigation system and standard CT-guided technique