Применение комплекса нагрузочных проб у пациентов с постоянной электрокардиостимуляцией при холтеровском суточном мониторировании
Воронина Наталья Владимировна, Viktoriia I. Skidan, Чирикова Ирина Александровна et al. · Dalʹnevostočnyj medicinskij žurnal · 2011
The study included 47 patients with permanent pacemaker, middle age 65,3±10,1, with complaints: weakness, dyspnea, fatigue, pain and discomfort in the fi eld of heart, dizziness, heart work impairment. Pacemakers in WI mode have been implanted to 23 (49%) patients, the two-chamber DDD/DDDR mode has been placed to 20 (42,5%), 4 (8,5%) patients had AAI modes. 24 hours Holter monitoring was performed under basic condition and against a complex of loading tests. As exercise stresses were used «Ladder test», the test with a strain of muscles of the top shoulder girdle. As psychomedical tests were used: the test of mirror drawing and the test «КОТ» V.N. Buzin. All taped electro-physiological deviations have been divided into fi ve groups: arrhythmias, conduction disturbances, pacemaker malfunction, especially with a pacemaker, ST segment depression on the spontaneous rhythm of the heart. All identifi ed deviations are presented in Table № 1. Signifi cantly more often such variations as: ventricular premature beats, pacemaker malfunction (stimulation, sensitivity, timing), confl uent complexes and pseudo-confl uent complexes, ST segment depression on the spontaneous rhythm of the heart were registered. Complex of loading tests during Holter monitoring in patients with permanent pacemaker help in timely diagnostics of the latent failure. Hence, complex of loading tests can be used for the patients under study.
Cited by 0 works.