高良姜素-高锰酸钾-多聚磷酸化学发光法测定牛血清蛋白
Li Li, 周敏, 马永钧 et al. · 2008
A novel flow-injection chemiluminescence (FI-CL) method for the determination of bovine serum albumin(BSA) is presented, based on the chemiluminescence reaction of BSA in the galangin-potassium permanganate-polyphosphoric acid system. Under the optimum conditions, the obtained linear range is 0.08~100 μg/mL, with a detection limit of 0.01 μg/mL(3σ).The regression equation is I(subscript CL)=220.448+7.068C(r=0.9997, n=9), and the relative standard deviation(RSD) for the determination of 1.0 μg/mL BSA is 1.5%(n=11). The proposed method is simple, rapid and high-sensitive. Two synthetic samples are determined selectively with recoveries in the range from 98.5 % to 100.2 %.
Cited by 0 works.