Повышение достоверности определения коэффициента открытой пористости газоволюметрическим пикнометром «Поромер»
Дахнов Алексей Владимирович, Крюкова Ирина Борисовна · Вести газовой науки · 2014
The gas-volumetric densimeter «Poromer» is designed for express evaluation of the open poro sity coefficient, mineralogical and volumetric density of specimens with correct geometric shape. The comparisons of the results of open porosity coefficient determination, defined by means of hydrostatic weighing (Preobrazhensky method) and the gas-volumetric method showed deviations in the results received, the reasons for which are explained by not only increased gas capability to penetrate to the porous space of the research specimen. One of the factors which can cause ‘apparent’ overestimation of the values of the open porosity coefficient defined with the Preobrazhensky method, as compared to the gas-volumetric method, is the accuracy of determination of the external volume for the specimen under study. In the gas-volumetric method the volume is calculated on the basis of the geometric sizes of the specimen. The calculation accuracy is defined by the degree of deviation of the real specimen shape from the ideal one. In the hydrostatic weighing method the specimen shape does not influence the reliability of determination of the external volume. Comparisons showed the necessity of introduction of an allowance in the results of determination of the open porosity coefficient received using the gas-volumetric densimeter «Poromer».
Cited by 0 works.