Микроколлапсы передней камеры глаза при ультразвуковой факоэмульсификации
Азнабаев Булат Маратович, Мухамадеев Тимур Рафаэльевич, Д. Р. Бикчураев et al. · Медицинский вестник Башкортостана · 2014
A new hydrodynamic scheme for the anterior chamber microcollapses compensation during ultrasound phacoemulsification was created. The scheme consists of a new suction pump and an aspiration line. An experimental and clinical estimation was conducted. The experiments proved that the amplitude of the postocclusion surge and the time required for recovery of intraocular pressure after its occurrence was less when the new hydrodynamic scheme was applied during the coaxial and microcoaxial phacoemulsification. The smallest morphological changes of experimental animals’ corneas were detected when the new hydrodynamic scheme was used. Clinical research showed that ultrasound phacoemulsification with the new hydrodynamic scheme is clinically safe and does not differ by a character and a frequency of complications from foreign analogs.
Cited by 0 works.