SENSE: Sensitivity encoding for fast MRI
Klaas P. Pruessmann, Markus Weiger, Markus Beat Scheidegger et al. · Magnetic Resonance in Medicine · 1999
New theoretical and practical concepts are presented for considerably enhancing the performance of magnetic resonance imaging (MRI) by means of arrays of multiple receiver coils. Sensitivity encoding (SENSE) is based on the fact that receiver sensitivity generally has an encoding effect complementary to Fourier preparation by linear field gradients. Thus, by using multiple receiver coils in parallel scan time in Fourier imaging can be considerably reduced. The problem of image reconstruction from sensitivity encoded data is formulated in a general fashion and solved for arbitrary coil configurations and k-space sampling patterns. Special attention is given to the currently most practical case, namely, sampling a common Cartesian grid with reduced density. For this case the feasibility of the proposed methods was verified both in vitro and in vivo. Scan time was reduced to one-half using a two-coil array in brain imaging. With an array of five coils double-oblique heart images were obtained in one-third of conventional scan time. Magn Reson Med 42:952-962, 1999.
Cited by 6,251 works. doi:10.1002/(sici)1522-2594(199911)42:5<952::aid-mrm16>3.0.co;2-s Open-access PDF
Authors
References
- Image Formation by Induced Local Interactions: Examples Employing Nuclear Magnetic Resonance
- The NMR phased array
- Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with radiofrequency coil arrays
- Fast MRI data acquisition using multiple detectors
- Fast imaging using subencoding data sets from multiple detectors
- A decoupled coil detector array for fast image acquisition in magnetic resonance imaging
- Correction for intensity falloff in surface coil magnetic resonance imaging
- An algorithm for compensation of surface-coil images for sensitivity of the surface coil
- Fast imaging method using multiple receiver coils with subencoding data set