ニワトリ胚体動計測のための任意酸素濃度制御システムの開発

晃希 天満, Asahi Yokota, 裕弥 千葉 et al. · IIAE Annual Conference · 2018

The purpose of this study is to develop a control system of an arbitrary oxygen (O2) concentration with quick responses and high accurate control.We used Arduino and Processing to create a system. Both of them are very similar to the C language and the knowledge of it learned so far can be utilized.In the control of the electromagnetic valve, a method is adopted in which the setting of the pulse width is determined for each target oxygen concentration.As a result, we could control the average arrival time and average recovery time within 1 minute and control the variation to within ± 0.1%.In the future research, we need to more data of the characteristic body movement patterns in the hypoxic environment by using our O2-concentration control system.

Cited by 0 works. doi:10.12792/iiae2018.012 Open-access PDF