Волоконный многоканальный гетеродинный интерферометр для исследования свойств импульсной плазмы
K S Lukyanov, К.Л. Губский, А.А. Ястребцев et al. · Письма в журнал технической физики · 2022
A fiber-based four-channel heterodyne interferometer has been developed to study the properties of pulsed plasma. The accuracy of phase measurements was 0.03 rad, which corresponds to an error in measuring the concentration of free electrons of ±1. × 1019 m-2. Measurements of the electron density and plasma velocity of the plasma accelerator have been carried out.
Cited by 0 works. doi:10.21883/pjtf.2023.01.54051.19354 Open-access PDF
Authors
References
- Digital I/Q demodulator
- A fiber–optic interferometer for in situ measurements of plasma number density in pulsed-power applications
- Interferometric investigation of a cablegun plasma injector
- Quadrature laser interferometry in the pulsed plasma diagnostic
- Two-Dimensional Phase Unwrapping via Balanced Spanning Forests