Экспериментальные исследования прочности бедренного компонента нового эндопротеза тазобедренного сустава

Nurlan Batpenov, Rolf Franke, У. Клеминг et al. · CyberLeninK (CyberLeninka) · 2013

Aim: to examine the strength and resistance to dynamic loads of the new femoral component of hip endoprosthesis. Materials and methods. Dynamic tests of left stem of cementless fixation prosthesis KAZ NIITO with diameter of 9 mm was carried out at laboratory of LLP “IMA research materials and application techniques” in Dresden (Germany). results: Experimental studies have shown that the strength of the new hip endoprosthesis is in line with ISO-Standard 7206-6 and can withstand maximum weight of 5340 N and more than 10 million cycles without damage, and can be used in clinical practice.

Cited by 0 works.