動画対応フラットパネルディテクタによる肺機能画像診断法 : 肺シンチグラフィ所見との比較

利恵 田中, 茂 真田, 政樹 藤村 et al. · Kanazawa University Repository for Academic Resources (DSpace) (Kanazawa University) · 2009

Pulmonary ventilation and circulation dynamics are reflected on dynamic chest radiographs as changes in X-ray translucency,i.e., pixel values. The present study was performed to develop a pulmonary functional evaluation method based on the changes in pixel value, and to investigate the clinical usefulness of our method. Sequential chest radiographs of 20 subjects (abnormal,n=12; normal,n=8) during respiration were obtained with a dynamic flat-panel detector (FPD) system. The average pixel value in each local area was measured tracking the same area. To facilitate visual evaluation, the results were mapped on the original image using a grayscale in which small changes were shown in black and large changes were shown in white. In our clinical evaluation in comparison with a pulmonary scintigraphy, pulmonary ventilation disorder was indicated as a reduction of changes in pixel values. In many patients, there was a correlation between our result and a pulmonary scintigraphy (0.7<r, 4 cases; 0.4<r<or=0.7, 6 cases; 0.2<r<or=0.4, 1 case; 0<r

Cited by 0 works. doi:10.24517/00014723 Open-access PDF