ポアソン過程よりみたクロマト分離の移動単位数

久弥 田中, 博幸 川崎, 辰美 山本 · Institutional Repositories DataBase (IRDB) · 1993

As a separation determining factor, the number of transfer unit N of chromatographic column was derived from an Poisson migration process of rate parameter χ = 1/τ_R at moving phase convection time t_c. It was obtained from the elution time distribution density function f(t) with the mean t_E and variance σ^2 for each component as (t_E-t_c)^2/σ^2. Calibrating the end effects of observed t_c, t_E and σ^2 of PEG system, the NTU could be plotted against t_c on several straight lines being determined by molecular weight. The degree of separation previously defined by (t_ -t_ )/SQRT(σ^2_1+σ^2_2) was linearly depended on the differences of square root N.

Cited by 0 works. doi:10.15099/00004457